Medicine stability testing chamber (second generation)

2024-07-23

Medicine stability testing chamber   (Second generation)

Scope of use:

Use scientific methods to create a temperature, humidity for drug failure evaluation, which is suitable for pharmaceutical companies to conduct experiments on factors affecting drugs and new drugs. 


Features:

Imported high-quality components are used to ensure long-term continuous operation and stable and reliable performance.

High-resolution dedicated touch screen temperature and humidity controller, excellent human-machine interface.

Standard three-level permissions and audit tracking function ensure data authenticity and consistency.

Imported high-precision humidity sensor with small drift and no maintenance required.

The standard micro printer can print temperature and humidity curves and data in real time.

RS485 interface is standard, which can monitor single or multiple instruments through computer.

Standard mobile phone SMS alarm, using Internet of Things technology, no SIM card required.

The visual observation window of the main door is equipped with a key lock to prevent unauthorized personnel from opening the door and affecting the test.

Multiple groups of imported fully enclosed heat-resistant compressors, low noise and long life.

Meet the relevant test requirements of FDA, ICH, JJF-2019 and national pharmacopoeia.

The circulating water system greatly delays the problem of bacteria growth in the water system.

It can meet the low temperature test requirements of 40℃/20%RH and 25℃/40%RH for special drugs such as large infusion .

Manufacturing execution standard (technical conditions): refer to GB/T 10586-2006.


Technical Parameters:

Model: LHH-150SD-2T,  LHH-250SD-2T,  LHH-500SD-2TLHH-800SD-2T,  LHH-1000SD-2T 

Volume (L): 150, 250, 500, 800, 1000

Temperature range: 0-65℃

Temperature fluctuation: ≤±0.5 

Temperature deviation: ≤±1 

Humidity range: 15-95%RH

Humidity deviation: ≤±3% RH

Humidification method: electric steam heating (external humidification, more energy-saving, avoids secondary pollution, and is more convenient for maintenance)

Working environment temperature: +5-40℃

Liner material: mirror stainless steel

Shell material: steel plate sprayed

Safety devices: independent overload protection, independent studio over-temperature protection, compressor overload protection, overpressure protection, water blockage protection, water shortage protection

Working power supply: AC220V 50HZ

Installation power (W): 1200, 1400, 2000, 2000, 2200

Plates: 2, 3, 4, 4, 4




Hotline: +86 400 887 8805

Company website: www.qinfutech.com 

Email: qinfutech@126.com



Beijing Qinfu Technology Co., Ltd. (North China): Manager Bao: 18514485008