Scope of use:
A temperature and humidity environment required for drug failure evaluation is created using scientific methods, which is suitable for pharmaceutical companies to conduct experiments on influencing factors of drugs and new drugs.
Features:
1. Use imported high-quality components to ensure long-term continuous operation and stable and reliable performance.
2. Original imported touch screen temperature and humidity controller, excellent human-machine interface.
3. Imported high-precision humidity sensor with small drift and no maintenance required.
4. The standard micro printer can print temperature and humidity curves and data in real time.
5. SD card is provided as standard, which can export the electronic record data in the controller.
6. RS485 interface is standard, which can be connected to the Internet and monitor single or multiple instruments through a computer.
7. Mobile SMS alarm is standard, and one mobile phone card can manage multiple devices.
8. The main door is locked with a key to prevent unauthorized personnel from opening the door and affecting the test.
9. Imported fully enclosed heat-resistant compressor, low noise, long life (one for backup and one for use).
10. Meet the accelerated, long-term, intermediate and high temperature and high humidity tests required by FDA, ICH and national pharmacopoeia.
11. The circulating water supply system can greatly delay the problem of bacteria growth in the water system.
12. Meet the low humidity test of 40℃, 20%RH and 25℃, 40%RH for special drugs such as large infusion.
13. Provide IQ, OQ and PQ verification data that meet international standards.
14. Manufacturing implementation standards (technical conditions): refer to GB10586-2006.
Technical Parameters
Model | QF-150SD-2T | QF-250SD-2T | QF-400SD-2T | QF-500SD-2T | QF-800SD-2T | QF-1000SD-2T | QF-2000SD-2T |
Volume (approx.) | 150L | 250L | 400L | 500L | 800L | 1000L | 2000L |
Temperature range | 10~65℃ |
Temperature fluctuation | ≤±0.5℃ |
Temperature deviation | ≤±1℃ (uniformity ≤2℃) |
Humidity range | 15~95%R.H |
Humidity deviation | ≤±3%RH (uniformity ≤5%RH) |
Humidification method | Electric steam humidifier (external humidification, more energy-saving; avoids secondary pollution and is more convenient for maintenance)
|
Working temperature | +5~40℃ |
Liner material | Mirror 304 stainless steel |
Shell material | Cold rolled steel plate spraying |
Safety devices | Power overload protection, independent studio over-temperature protection, compressor overload, overpressure protection, water blockage, water shortage protection |
Working power supply | AC220V±10% 50HZ |
Installed power | 1.2KW | 1.4KW | 2KW | 2KW | 2KW | 2.2KW | 3KW |
Carrier Plate | 2 layer | 4th floor | 4th floor | 4th floor | 4th floor | 5th floor | Level 6 |
Note: Dimensions are D x W x H(mm)