Scope of use:
1. Use scientific methods to create a long-term temperature and humidity environment required for drug failure evaluation.
2. It is convenient for storing large quantities of drugs and conducting long-term moisture resistance cycle tests.
Features:
1. Use imported high-quality components to ensure long-term continuous operation and stable and reliable performance.
2. Original imported special touch screen temperature and humidity controller, with clear and intuitive display, stable control and easy operation.
3. Configure three-level permissions, audit tracking, and electronic signatures.
4. Imported high-precision humidity sensor with high sensitivity, small drift and no maintenance required.
5. Equipped with a micro printer to print temperature and humidity curves and data in real time.
6. Optional monitoring and recording system that complies with FDA standards, double recording makes data safer.
7. The standard RS485 interface can be used to control one or more instruments through a computer and to record and back up control data in real time.
8. Standard 4G temperature and humidity deviation and power outage SMS alarm function. One mobile phone card can know the operating status of multiple instruments.
9. Equipped with a door key lock and anti-reverse locking device to prevent irrelevant personnel from affecting the experiment, and people can exit the laboratory safely.
10. The automatic circulating water supply system solves the problems of frequent water addition, water tank cleaning and no drainage pipes, greatly delaying the problem of bacteria growth in the water system.
11. Imported fully enclosed heat-resistant compressor adopts environmentally friendly refrigeration, low noise, and long life. The outdoor unit mode can be designed according to the on-site conditions to save energy and reduce emissions.
12. Meet the accelerated, long-term, intermediate and high temperature and high humidity tests in the relevant requirements of FDA, ICH, JJF-2019 and the National Pharmacopoeia.
13. Provide IQ, PQ and OQ verification data that meet international standards.
14.Manufacturing implementation standards (technical conditions): refer to GB/T 10586-2006.
Technical Parameters:
model | LHH-9000SDP/FD | LHH-15000SDP/FD | LHH-30000SDP/FD | LHH-40000SDP/FD | LHH-***SDP/FD |
Volume (M3) | 9 | 15 | 30 | 40 | Customized 5-500 cubic meters of content |
temperature range | 20~50℃ |
Temperature fluctuation | ≤±0.5℃ |
Temperature deviation | ≤±1℃ (uniformity ≤2℃) |
Humidity range | 30~90%RH (external humidification, more energy-saving; avoid secondary pollution and more convenient maintenance) |
Humidity deviation | ≤±3%RH (uniformity ≤5%RH) |
Humidification method | Electric steam humidifier |
Working temperature | +5~40℃ |
Liner material | 304 stainless steel |
shell material | Cold rolled steel plate spraying |
safety devices | Power overload, independent over-temperature, compressor overload, over-pressure, water blockage, water shortage protection |
Working power supply | AC220V±10% 50HZ | AC220V±10% 50HZ | AC220V or 380V±10% 50HZ |
Installation power (W) | 3500 | 4500 | 5500 | 4500 | Determined by user volume |
Carrier | 4pcs/4layers | 4pcs/4layers | 8pcs/4 layers | 12pcs/4 layers | Determined by user volume |
Internal dimensions (mm)* | 2000X2100X2200 | 2000X3400X2200 | 3600X3780X2200 | 3600X5050X2200 | Determined according to the user's site size |
Dimensions (mm)* | 2200X2300X2400 | 2200X3600X2400 | 3800X3980X2400 | 3800X5250X2400 | Determined according to the user's site size |
Remark | The operating power is the actual measured data of 40℃ 75% no-load with one compressor turned on; the cargo rack is available in nickel-plated, stainless steel and other options |
Note: Dimensions are D x W x H(mm)