Medicine High-light Exposure Testing Chamber


  • 价格:Negotiated
  • 编号:LHH-GD-T
  • 产地: Shanghai Qinfu
Details
Specification
Parameter

Scope of use:


It is applicable to the influencing factor test and strong light irradiation stability test of drugs to determine the validity of the test samples.



Features:


1. Use imported high-quality components to ensure long-term continuous operation and stable and reliable performance.

2. Intelligent temperature controller imported from Japan, stable control, easy operation and long service life.

3. The standard micro printer can print temperature, illumination, near-UV curves and data in real time for traceability.

4. Standard RS485 interface can be connected to the Internet to monitor single or multiple instruments through a computer.

5. The main door is locked with a key to prevent unauthorized personnel from opening the door and affecting the test.

6. The illumination adopts LED white light source, and the irradiance adopts LED ultraviolet light source. The illumination is automatically controlled in a closed loop, with good uniformity and large effective area.

7. Imported fully enclosed heat-resistant compressor, R134a environmentally friendly refrigeration, low noise and long life.

8. Standard independent temperature limit alarm system, the instrument will stop running when the temperature exceeds the limit.

9. Meet ICH:Q1B guidelines: total illumination ≥1.2×106Lux.hr, near-ultraviolet energy ≥200W.hr/m2.

10. Meet the strong light irradiation test of 4500±500Lux in the 2015 edition of the chemical drug stability guidelines.

11. Provide IQ, OQ and PQ verification data that meet international standards.



Technical Parameters:


Product number

Drug strong light irradiation test chamber (light + UV)

model

LHH-150GD-T(-F)

LHH-250GD-T(-F)

LHH-500GD-T(-F)

Volume (L)

150

250

500

temperature range

10~65℃(Full light 15~65℃) (-F2-8℃)

Temperature fluctuation

≤±0.5℃

Temperature deviation

Same plane≤±2℃

Illumination range

10-10000LUX

Illuminance deviation

±500LUX

Near UV illumination range

80~200μw/cm2(UVA)

Near ultraviolet wavelength range

320~400nm

Working temperature

+5~40℃

Liner material

Mirror stainless steel

shell material

Cold rolled steel plate spraying

safety devices

Power overload, independent overtemperature, compressor overload, overpressure

Working power supply

AC220V±10% 50HZ

Installation power (W)

600

800

1200

light source

1 story

2 layer

3 layers (can be increased to 4 layers)

Internal dimensions (mm)

500×530×600

500×530×950

500×630×1600

Dimensions (mm)

820×690×1410

820×690×1750

820×1190×1890

Note: Each model can be customized with a 2~8℃ low temperature light test chamber (-F).



Note: Dimensions are D x W x H(mm)