Medicine Refrigerating Testing Chamber

2024-11-21

Medicine Refrigerating Testing Chamber

Scope of use:

It is suitable for various pharmaceutical companies, scientific research institutes, etc. to conduct cold storage tests on samples to determine the validity of the test samples.


Features:

1. Use imported high-quality components to ensure long-term continuous operation and stable and reliable performance.

2. Original imported intelligent temperature controller with high control precision, easy operation and long service life.

3. Equipped with a micro printer to print temperature curves and data in real time for traceability.

4. Equipped with an independent temperature limit alarm system, the instrument will stop running when the temperature exceeds the limit.

5. The main door is locked with a key to prevent unauthorized personnel from opening the door and affecting the test.

6. Imported fully enclosed compressor, environmentally friendly refrigeration, low noise and long life.

7. The airflow in the box adopts convection circulation to make the temperature inside the box uniform.

8. Optional RS485 interface can be connected to the Internet to monitor single or multiple instruments through a computer.

9. Optional mobile phone SMS alarm, one mobile phone card can manage multiple devices.

10. Meet the 2~8℃ low temperature refrigeration test required by FDA, ICH and national pharmacopoeia.

11. Stable and reliable performance, simple operation, low power consumption and low noise.

12. Manufacturing implementation standards (technical conditions): refer to GB/T 10589-2008


Technical Parameters


Product name

Medicine Refrigerating Testing Chamber 

Model

LHH-150MP

LHH-250MP

LHH-500MP

LHH-800MP

LHH-1000MP

Volume (L)

150

250

500

800

1000

Temperature range

2~8℃

Temperature fluctuation

≤±0.5℃

Temperature deviation

≤±2℃

Working temperature

+5~30℃

Liner material

Mirror stainless steel

Shell material

Cold rolled steel plate spraying

Safety devices

Power overload protection, independent studio over-temperature protection, compressor overload, overpressure protection

Working power supply

AC220V±10% 50HZ

Note: * Dimensions are D x W x H (mm)



Hotline: +86 400 887 8805

Company website: www.qinfutech.com / www.qinfukj.com

Email: qinfutech@126.com

Beijing Qinfu Technology Co., Ltd. (North China): Manager Bao: 18514485008

Shanghai Qinfu Technology Co., Ltd. (East China): Manager Qian: 15021305696

Guangzhou Qinfu Technology Co., Ltd. (South China): Manager Zhao: 18796797820

Chongqing Qinfu Technology Co., Ltd. (Southwest China): Manager Ran: 15000109726

Hunan Qinfu Technology Co., Ltd. (Central China): Manager Yang: 15021307892